Micro Photonics Inc.

MF-ChemiBIS Gel Imager System

DNR's breakthrough MF-ChemiBIS family is the most advanced image analysis system in the market for 1D and 2D fluorescent, chemiluminescent, and colorimetric applications.

With advanced robotic camera, controlled by the software and motorizes up and down automatically, DNR's new line of imaging systems outperforms the standard zoom lens by a wider field of view, easier and faster operation, and better sensitivity, which all have a direct impact on the results, especially in faint samples. A vibration free technology of a Deep Liquid Cooling Technology cools the camera by liquid to a temperature down to -70ºC. DNR is the first to implement a liquid cooling technology in imaging systems. The technology outperforms all other methods currently in the market and results in better accuracy, longer exposure time (up to 24 hours), minimal dark noise and full stability of the system.

The MF-ChemiBIS smart dark chamber is fully automated (camera, iris, focus, filter wheel) and includes DNR's unique Automatic Drawer Detection Technology of the three different drawers (UV, WL, ECL) supplied with the system for an optimal match with the light source and sample.

The MF-ChemiBIS family standard with DNR's acquisition software - GelCapture - for complete, automatic control of configuring, capturing and saving sample for analysis in a minimum amount of time.

 

 

  • Breakthrough Robotic camera- allows placement very close to the sample, resulting in very high sensitivity and superior image quality under a wide range of conditions.
  • High resolution- selectable scientific camera (2.0 Mpixel, 3.2 Mpixel) gives maximum band separation and best quality images.
  • Unrivalled sensitivity- high speed super bright lens collects maximum light from the sample while maintaining crisp, clear images with minimal geometric and light distortion.
  • Novel Deep Liquid Cooling System- cools the camera to minimum of -70°C for better accuracy, longer exposure time and minimal dark noise.
  • Illumination versatility- 10 types of illumination sources to suite a wide range of samples and applications.
  • Fully automated and motorized- permits simultaneous control of a number of settings such as: zoom, focus, gain and more, for easier image capture and faster results

System
MF-ChemiBIS 2.0
MF-ChemiBIS 3.2
Camera & Optics
CCD resolution
1,640 x 1,214; 2.0 Mpixel
2,184 x 1,510; 3.2 Mpixel
Gradation
16 bit (65,536 gray levels)
Dynamic range
Approx. 5  orders of magnitude
Interface
USB 2.0
Binning
1 x 1, 2 x 2, 3 x 3, 4 x 4, 5 x 5, 6 x 6, 7 x 7, 8 x 8
Illumination & Darkroom
Dark chamber
DNR's Smart Dark Chamber
Drawer
3 interchangeable drawers : UV, ECL, optional: WL
Emission filters
5-position filter wheel with 3 standard filters (included)
Low UV 520 nm
High UV 580nm
Red 600 nm
Illumination modes
Trans-UV, Trans-WL, Epi-WL
 
Optional: Epi-UV, QuattroLED illumination kit fro advanced fluorescnece appliations
Excitation sources
White light 
UV: 312 nm
 
Optional: 254 nm  or 365 nm
 
Optional: dual UV source
QuattroLED Iillumination system (optional)
470 nm, 525 nm, 620 nm, 740 nm
Software
Image capture
GelCapture, lifetime free upgrade
Image analysis
GelQuant Pro for 1D analysis
 
Optional 1D analysis software: Phoretix 1D, Phoretix 1D Pro(Nonlinear Dynamics)
 
Optional 2D analysis software: Prodigy Same Spot(Nonlinear Dynamics)
Power
100 or 230 VAC 50-60 Hz
Band cutting illuminator
Designed specifically to cut bands

 

  • Fluorescence - Ethidium bromide, SYPRO™ Orange, SYPRO™ Red, Cy ™ 2, Cy ™ 3, Pro-Q™ Diamond, Deep Purple™, Rhodamine Red, SYPRO™ Ruby, SYBR™ Green, GelStar™, SYBR™ Gold, Fluorescein, Texas Red™, GFP, and more
  • Colorimetric- Silver Stain, Coomassie Brilliant Blue, film, and more
  • Chemiluminecsence - ECL™, ECL-Plus™, Lumi Light Plus™, Super Signal™, CDP Star™, CSPD™
  • Other - TLC plates, spot and slot blots, autoradiographs, microplate, macroarrays, membranes, and more